1-4 Safety Precautions
10.
Do not allow the patient to contact the live parts of the ultrasound
system or other devices, e.g. signal I / O ports. Electric shock may
occur.
11.
Do not use an aftermarket probe other than those specified by
Mindray. The probes may damage the system causing a profound
failure, e.g. a fire in the worst case.
12.
Do not subject the probes to knocks or drops. Use of a defective
probe may cause an electric shock.
13.
Do not open the covers and front panel of the system. Short circuit
or electric shock may result when the system hardware is exposed
and powered on.
14.
Do not use this system when any digital device such as a high-
frequency electrotome, high-frequency therapeutic device or
defibrillator is applied already. Otherwise, there is a risk of electric
shock to the patient.
15.
Only use the ECG leads provided with the physiology module;
otherwise, electric shock may be resulted.
16.
When moving the system, you should hold the handle; otherwise,
damage may be resulted by abnormal force. Do not push the
system from the left/right side; otherwise, it may be toppled over.
17.
The auxiliary power output outlet in the system is used to supply
power for the recommended peripheral devices. Do not connect
other devices to the outlet, otherwise, the rated output power
limitation may be exceeded and failure may be resulted. Maximum
output power of the outlet is 240VA.
18.
Accessory equipment (analog or digital) connected to the
ultrasound system must comply with the relevant IEC standards
(e.g., IEC 60950 information technology equipment safety standard
and IEC 60601-1 medical equipment standard).Furthermore, all
configurations must comply with the standard IEC 60601-1-2:2014. It
is the responsibility of the person, who connects additional
equipment to the signal input or output ports and configures a
medical system, to verify that the system complies with the
requirements of IEC 60601-1-2:2014. If you have any questions
regarding these requirements, please consult your sales
representative.
19.
Prolonged and repeated use of keyboards may result in hand or arm
nerve disorders for some individuals. Observe the local safety or
health regulations concerning the use of keyboards.
20.
When using intra-cavity probes, do not activate the probe outside
the patients body.
21.
DO NOT touch the Signal I/O ports if in contact with the patient;
otherwise patient injury may result.