EasyManuals Logo

Mindray DC-85 User Manual

Mindray DC-85
383 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Page #350 background imageLoading...
Page #350 background image
B-2 Ultrasound Gel Warmer
No.
Name
Function
5.
Bottom cover
/
B.3 Specifications
Power supply
Voltage
12V
Power consumption
12W
Environmental conditions
Operating conditions
Storage and transportation conditions
Ambient temperature
5℃~40
-20℃~55
Relative humidity
35%85% (no condensation)
30%85% (no condensation)
Atmospheric pressure
700hPa1060hPa
700hPa1060hPa
B.4 Function and Requirement
The ultrasound gel warmer can heat the ultrasound gel to about 40
1 at most.
When the gel is put inside the warmer, the time it takes to heat from environmental condition of 18 to
28 should be no more than 0.5 hours.
The ultrasound gel warmer can work continuously over 12 hours.
NOTE:
1.
When the environment temperature is higher than the requirement temperature of
the warmer, the warmer does not function.
2.
The warmer can heat only one bottle of gel at one time.