Overview 
weighting circuit. 
If the transmission pulse width is different, the transmission energy is different. For the 
elements involving in the transmission, if the transmission pulses with different width are 
used to excite the elements, the transmission-weighting effect can be obtained. 
For single pulse transmission, since the adjustment of transmission pulse width is limited, 
the transmission weighting is only used for the transducers with low frequencies. For 
multiple   pulse   transmission,   since   the   central   frequency   is   determined   by   the   pulse 
repetition period, the high level width of transmission pulse can be adjusted in a bigger 
range, and the transmission-weighting effect is better. The transmission weighting is used 
for all transducers. 
2.1.1.3Reception circuit
The reception circuit executes the  selection on the reception elements, element echo 
amplification and AD conversion. The circuit is comprised of high voltage isolation and 
reception selection circuit, voltage-controlled gain amplifier, anti-aliasing filtering, AD and 
ATGC. The diagram of the reception circuit is shown as follows: 
3-6 DP-3300/DP-3200 Service Manual(V1.1)