Electrical Safety Inspection
A-9
ELECTRICAL SAFETY INSPECTION
6- Patient Leakage Current
OVERVIEW
Patient leakage currents are measured between a selected applied part and mains earth. All
measurementsmayhaveeitheratrueRMSoraDConlyresponse.
TEST PROCEDURE
Prepare
PerformacalibrationfromtheMainsonAppliedPartmenu.
Thefollowingoutletconditionsapplywhenperformingthistest:
NormalPolarity ,EarthOpen,OutletONNormalPolarity ,OutletON
NormalPolarity ,L2Open,OutletONReversedPolarity ,OutletON
ReversedPolarity ,EarthOpen,OutletONReversedPolarity ,L2Open,OutletON
WARNING
If all of theapplied parts correspond to the instrument type, theapplied parts will betied
together and one reading will be taken. If any of the applied parts differ from the instrument
type,allappliedpartswillbetestedindividually,basedonthetypeofappliedpart.Thisapplies
to
AutoandStepmodesonly.
Perform the Test
1) FromtheMAINMENU,orwiththe outletunpowered,plugtheDUTinto the601PROfront
paneloutlet,andturnonthedevice.
2) Attachtheappliedpartstothe601PRO'sappliedpartterminals.
3) Pressshortcutkey6.ThePatientLeakagetestisdisplayed,andthetestbeginsimmediately.
4) Press APPLIED PART (SOFT KEY 4) at any time to select the desired applied part leakage
current.