MR6401 CPR Sensor Operator’s Manual
Physical
Specifications
Sensor size < 150mm × 63mm × 20m
Compression area < 92mm × 53mm × 19mm
Adhesive mount area ≤ 98mm × 45mm
Sensor weight ≤ 220g (including battery, without CPR sensor cable)
Measurement
Specifications
Compression depth Measurement range 0.0 to 8.0 cm
Effective range 1.5 to 8.0 cm
Accuracy ±0.5 cm or ±10%, whichever
is greater
Resolution 0.1 cm
Refreshing rate ≥ 0.5 Hz
Compression rate Measurement range 40 to 160 cpm
(compressions per minute)
Effective range 40 to 160 cpm
(compressions per minute)
Accuracy ±2 cpm (compressions per
minute)
Resolution 1 cpm
Refreshing rate ≥ 0.5 Hz
Interruption time Measurement range 0 to 300 s
Effective range 0 to 300 s
Resolution 1 s
Refreshing rate ≥ 0.5 Hz
EMC
The device meets the requirements of IEC 60601-1-2. For more information, see the operator’s manual
of the defibrillator or monitor used together with this device.
Symbol
Refer to instruction manual/booklet
Date of manufacture
Serial number
IP55
Dust-protected
Protected against water jets
DEFIBRILLATION-PROOF TYPE CF
APPLIED PART
Authorized representative in the
European Community
Not made with natural rubber latex.