Overview
6 Operator's Manual
Item Quantity Description
Carryingcase 1 /
Tipcover 1 /
Transducercoversuite 1 OnlysuppliedwithP8-2Ts/P8-2TU.
(IncludingNon-sterileTEEcover,gel-lled
syringeandbite-guard)
Tips:
Thetipcoverenclosesandprotectsthedistalend/scanheadoftheendoscopefrombeing
exposedtomechanicalstrainduringtransportationandstorage.
2.3.2 Transducer Inspection
ThefollowinginspectionsshouldbeperformedontheTEEtransducerafterunpackingthe
contents.
Performvisualandtactileinspection,see"3.1.1 VisualandTactileInspection".
Performtipdeectioninspection,see"3.1.2 TipDeectionControl".
Performbrakeinspection,see "3.1.3 Brake Operation".
Performscanplanerotationinspection,see"3.1.4 ScanplaneRotationOperation"..
Performleakagetest,see"3.3 ElectricalSafety".
ContactMindrayoryourlocalrepresentativeimmediatelytoreportanydamageor
discrepancies.
WARNING
To avoid injury to the patient, if any irregularity, substandard function or unsafe
condition is observed or suspected, the TEE transducer should not be used.
2.4 Structure
TEEtransducerisaphasedarraytransducer,mountedinasealedtipattheendofa
gastroscope.Thetransducerisintendedforimagingoftheheartthroughtheesophagus
andthestomach.Thearraycanberotated0°–180°.