EasyManua.ls Logo

MITECH MH320 - 4 Preparation & Testing; 4.1 Preparation & Inspection before Testing; 4.1.1 Preparation of Sample Surface; 4.2 Testing Program

Default Icon
27 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
7
mvip@mitech-ndt.com
Table 3-1
No.
Type of impact
device
Hardness value of Leeb
standard hardness block
Error of displayed
value
Repeatability
1
D
760±30HLD
530±40HLD
±6 HLD
±10 HLD
6 HLD
10 HLD
2
DC
760±30HLDC
530±40HLDC
±6 HLDC
±10 HLDC
6 HLD
10 HLD
3
DL
878±30HLDL
736±40HLDL
±12 HLDL
12 HLDL
4
D+15
766±30HLD+15
544±40HLD+15
±12 HLD+15
12 HLD+15
5
G
590±40HLG
500±40HLG
±12 HLG
12 HLG
6
E
725±30HLE
508±40HLE
±12 HLE
12 HLE
7
C
822±30HLC
590±40HLC
±12 HLC
12 HLC
Measuring range:HLD(170~960)HLD
Measuring direction:360°
Hardness Scale:HL、HB、HRB、HRC、HRA、HV、HS
Display:dot matrix LCD,128×64 dots
Data memory: 500 groups max.(relative to impact times 32~1)
Printing paper: width is (57.5±0.5)mm, diameter is 30mm
Battery pack: 6V NI-MH
Battery charger: 9V/500mA
Continuous working period:about 150 hours(With backlight off, no printing)
Communication interface:USB1.1
sale@ponpe.com

Related product manuals