7
mvip@mitech-ndt.com
Table 3-1
No.
Type of impact
device
Hardness value of Leeb
standard hardness block
Error of displayed
value
Repeatability
1
D
760±30HLD
530±40HLD
±6 HLD
±10 HLD
6 HLD
10 HLD
2
DC
760±30HLDC
530±40HLDC
±6 HLDC
±10 HLDC
6 HLD
10 HLD
3
DL
878±30HLDL
736±40HLDL
±12 HLDL
12 HLDL
4
D+15
766±30HLD+15
544±40HLD+15
±12 HLD+15
12 HLD+15
5
G
590±40HLG
500±40HLG
±12 HLG
12 HLG
6
E
725±30HLE
508±40HLE
±12 HLE
12 HLE
7
C
822±30HLC
590±40HLC
±12 HLC
12 HLC
Measuring rangeHLD170960HLD
Measuring direction360°
Hardness ScaleHLHBHRBHRCHRAHVHS
Displaydot matrix LCD128×64 dots
Data memory 500 groups max.relative to impact times 321
Printing paper: width is 57.5±0.5mm, diameter is 30mm
Battery pack: 6V NI-MH
Battery charger: 9V/500mA
Continuous working periodabout 150 hoursWith backlight off, no printing
Communication interfaceUSB1.1