8
2 Preparations before Use
No. 99MAH054A
Tips
When using this product in a posture where the contact point is facing downward to
horizontal, use the standard coil spring.
Store the removed coil spring to prevent loss.
2.2
Mounting to a Stand/Jig
2.2.1 Mounting by Stem
Install the stem to the holder of the stand when making measurement with this
product installed on a stand, etc.
NOTICE
Wheneverpossible,avoidxingthestemdirectlywithasetscrew,etc.
The plunger may not be able to move smoothly if the screw is tightened with a
tightening torque of 300 cN•m or more to secure the stem.
21
Stem
ø8*
+0.02
mm
+0.005
*ø9.52 mm (0.375 in)
ID-F0512ENX, ID-F0512ENXB,
ID-F0525ENX, ID-F0550ENX,
ID-F0550HENX
1 Install the stem to a holder such as a stand.
2 Tighten the holder screws.
Tips
Use a slotted holder with a *ø8 G7 (+0.005 to +0.02) mm hole for the holder/stand.
*ø9.52 mm (0.375 in): ID-F0512ENX, ID-F0512ENXB, ID-F0525ENX, ID-F0550ENX, ID-F0550HENX