71
No. 99MAH054A
7 Error Displays and Countermeasures
7
Error Displays and Countermeasures
No. Error name LCD display Cause Countermeasures
-
ABS Synthesis
Error
Last digit is E.
The sensor
signal cannot be
synthesized.
Although this may occur
during high-speed
plunger movement,
thereisnoe󰀨ecton
measurement.
Use the product as is.
* If it occurs while the
plunger is not moving,
the sensor may have
failed. Please contact
the agent where you
purchased the product
or a Mitutoyo sales/
service representative.
( "11O󰀨-SiteRepairs
(Subject to Charge)" on
page 87)
15
Low Battery
Voltage Error
Err 15
Cannot perform
measurement due
to a decrease in
supply voltage.
The supply voltage may
have dropped or become
unstable. Check the
supply voltage.
30
Display Value
OverowError
Err 30
The display value
exceeds the
displayable range.
When the display value
returns to the number of
displayable digits, the
error is automatically
cleared.
y Reset the resolution.
y Set the calculation
coe󰀩cientagain.
(
"5.1 Selecting
Parameter Items" on
page 37,
"5.5
Selecting Resolution"
on page 43,
"5.7
Setting Calculation
Function Selection and
CalculationCoe󰀩cient"
on page 48)