EasyManua.ls Logo

Mitutoyo ABSOLUTE ID-F0550HENX - Setting and Value Errors

Mitutoyo ABSOLUTE ID-F0550HENX
102 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
73
No. 99MAH054A
7 Error Displays and Countermeasures
No. Error name LCD display Cause Countermeasures
51
Calibration
Schedule
Warning Forced
Error
Err 51
A serial
communication
command [89]
(calibration
schedule warning
forced display) is
received.
Send a serial
communication
command [89]
(calibration schedule
warning forced display).
61
Set Value
Rewrite Error
Err 61
For some reason,
the set value has
been rewritten
since the previous
use.
Press the [F1] key to
return to measurement
mode, check various
settings, and reset where
needed.
62
Set Value
Storage Error
Err 62
Settings cannot be
saved.
The set values
cannot be read.
y After turning OFF the
power, turn ON the
power again, check
various set values,
and reset where
needed.
y If the same error
occurs even after the
power is turned on
again, the product
may be faulty. Please
contact the agent
where you purchased
the product or a
Mitutoyo sales/service
representative.
(
"11O󰀨-Site
Repairs (Subject
to Charge)" on
page 87)
y If the error occurs
frequently, the supply
voltage may be
unstable. Check the
supply voltage.

Table of Contents