4 Basic Operations
29
No. 99MBC109B
Tips
Each time you press [SEL], the blinking switches between Display A and Display B.
"4.1 Switching Between Display A and B (2-Axis Models Only)" (page 25)
2
Press [MODE].
»
Peak mode will be set.
3
Press [A_ZERO]/[ZERO] or [B_ZERO].
»
The peak value will be cleared (MAX = MIN = current value, TIR = 0).
Tips
If the same UNIT (counting method) has been assigned to more than 1 CEL, the peak
value of all of the CELs that are assigned the same UNIT (counting method) will be
cleared when the peak value of one of the CELs is cleared.
For details about the relationship between BANK, CEL, and UNIT, see "3.4.1 Internal
Structure of the Counter" (page 17).
4.6
Tolerance Value Setting
There are 2 settings for the tolerance value: 3-step and 5-step.
The tolerance values can be set separately for each CEL.
Tips
Set the value of parameter number 20 to 0 (3-step tolerance) or 1 (5-step tolerance) in
advance.
For details about I/O output, see
"5.5 I/O Connector Terminal Function" (page 42).
4.6.1 3-Step Tolerance Value Setting (3-Step Tolerance
Zone Selection)
With S1 and S4 set as the tolerance values, the 3-step tolerance judgment will
be performed as follows:
Judgment conditions
Tolerance judgment indicator
I/O output (PIN number)
Measurement result < S1 Amber indicator on AL1 (3) / BL1 (11)
S1 ≤ measurement result ≤ S4 Green indicator on AL3 (5) / BL3 (13)
S4 < measurement result Red indicator on AL5 (7) / BL5 (15)
This section explains how to set the 3-step tolerance value.
1
Press and hold [SEL]. (For 1-axis models, proceed to step
2
.)
»
The selected Display will blink.
Tips
Each time you press [SEL], the blinking switches between Display A and Display B.
"4.1 Switching Between Display A and B (2-Axis Models Only)" (page 25)