5 External Input/Output Function
44
No. 99MBC109B
5.5.2 Output Function
Output of total tolerance judgment result
On 2-axis models, you can output the Total Judgment result of both INPUT A
and INPUT B and the Total Judgment result of multiple Counters when multiple
Counters are connected.
ALLGO ALLGO ALLGO
EH EH EH
Judgment
Total
Judgment
indicator
External
output
ALL_GO
All Counters OK Green H
All or some
Counters NG
Amber L
Error Red L
Tips
To enable this function, select 0 or 1 as the set value of parameter number 20.
For the Total Judgment across multiple Counters, wire the number 8 pins (ALLGO: total
tolerance judgment result output) together.
For USB output, the Total Judgment indicator lights in amber in normal conditions and in
red when an error has occurred. Perform the total tolerance judgment with SENSORPAK
(optional software).
For BCD output, the Total Judgment indicator lights in amber in normal conditions and in
red when an error has occurred.
For details about the timing chart, see "5.5.4 Timing Chart" (page 46).
BCD output
Outputs measurement data in BCD format. Measurement data from INPUT A
and INPUT B is output simultaneously in 4-bit units beginning with the last digit.
An overview of the data format is as follows.
D8D4D3D 2D1
A/B_bit3
A/B_bit2
A/B_bit1
8x10
0
4x10
0
2x10
0
1x10
0
A/B_bit0
D8D7D6D5D4D3D2D1
12345601
-12345601
A/BSIGN
1x10
1
2x10
1
4x10
1
8x10
1
1x10
7
2x10
7
4x10
7
8x10
7
BANK
Bit0-3
SIGN L
+:H
-:L
PEAK1 PEAK2
PEAK1 PEAK2
NOM
L
L
MAX
H
L
MIN L H
TIR H H
BANK1:L
BANK2:H
LSD (least signicant digit)
Sign
Data output example
Not used Not used
MSD (Most signicant digit)