Compliance and EMC Declaration
Safety Standards: EN 60601-1: 2006+A1:2013, EN 60601-1-2:2007
Internally powered equipment (battery power)
Front Panel and Case Labelling
Home Healthcare Environment
Guidance and Manufacturers Declaration – Electromagnetic Emission
The Non-Contact Infrared Body Thermometer is intended for use in the electromagnetic environment specified below. The
user of the Non-Contact Infrared Body Thermometer should ensure that it is used in such an environment.
Electromagnetic Environment – Guidance
The Non-Contact Infrared Body Thermometer uses RF energy
only for its internal function. Therefore, its RF emissions are
very low and are not likely to cause any interference in
nearby electronic equipment.
The Non-Contact Infrared Body Thermometer is suitable for
use in all establishments, other than domestic and those
directly connected to the public low voltage power supply
network that supplies building used for domestic purposes.
Harmonic Emissions IEC
61000-3-2
Voltage
Fluctuations/Flicker
Emissions IEC 61000-3-3
Guidance and Manufacturers Declaration – Electromagnetic Immunity
The Non-contact Infrared Body Thermometer is intended for use in the electromagnetic environment specified below. The
user of Non-Contact Infrared Body Thermometer should ensure that it is used in such an environment.
Electromagnetic Environment – Guidance
Electrostatic discharge
(ESD) IEC 61000-4-2
Floors should be wood, concrete or
ceramic tile. If floors are covered with
synthetic material, the relative humidity
should be at least 30%
Electrical fast
transient/burst IEC
61000-4-4
±2 kV for power supply lines
±1 kV for input/output lines
Mains power quality should be that of a
typical commercial or hospital
environment.
±1 kV line(s) to line(s)
±2 kV line(s) to earth
Mains power quality should be that of a
typical commercial or hospital
environment.
Voltage dips, short
interruptions, and
voltage variations on
power supply input lines
IEC 61000-4-11
<5% U
T
(>95% dip in U
T
) for 0.5
cycle
40% U
T
(60% dip in U
T
) for 5 cycles
70% U
T
(30% dip in U
T
)for 25 cycles
<5% U
T
(>95% dip in U
T
) for 5 sec
Mains power quality should be that of a
typical commercial or hospital
environment. If the user of the Non-
Contact Infrared Body Thermometer
requires continued operation during
power mains interruptions, it is
recommended that the Non-Contact
Infrared Body Thermometer be powered
from an uninterruptible power supply or a
battery.
Power frequency
(50Hz/60Hz) magnetic
field IEC 61000-4-8
Power frequency magnetic fields should
be at levels characteristic of a typical
location in a typical commercial or
hospital environment.
Note: U
T
is the a.c. mains voltage prior to application of the test level.