37
Troubleshooting
Unit does not turn
on when button
is pushed.
Measuring
Error Symbol appears
when blood pressure
value displayed is
excessively low
or high.
Measuring Error
Symbol
Measuring Error
Symbol
Measuring Error
Symbol
SYMBOLS /
SYMPTOMS
CONDITIONS /
CAUSES
INDICATION /
CORRECTION
Worn-out batteries.
Battery polarities
have been positioned
incorrectly.
Cu has been placed
incorrectly.
Did you talk or move
during measurement?
Shaking of the wrist
with the cu on.
Ination pressure
exceeding 300 mmHg.
Can’t determine blood
pressure measurement
data.
Replace them with
2 new AAA (LR03)
alkaline batteries.
Re-insert the batteries in
the correct positions.
Wrap the cu
properly so that it is
positioned correctly.
Measure again. Keep
wrist steady during
measurement.
Check cu connection.
Measure again.
Switch the unit o,
then measure again.
Wrap the cu properly
and keep steady.
Measure again.
Note: If "EP" appears on the display, just return the device to your local
distributor or importer.
Air circuit abnormality.
Cu tube may not be
plugged into monitor
correctly.