47
Appendix
Guidance and manufacturers declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments listed
below, and should only be used in such environments:
Immunity test
IEC 60601 test
level
Electromagnetic
environment – guidance
Electrostatic
discharge (ESD)
IEC 61000-4-2
±6 kV
contact
±8 kV air
The relative humidity
should be at least 5%.
Compliance
level
Power
frequency
(50/60 Hz)
magnetic eld
IEC 61000-4-8
±8 kV
contact
±15 kV air
3 A/m
3 0A/m
Power frequency magnetic
elds should be at levels
characteristic of a typical
location in a typical
commercial or hospital
environment. Recommended
separation distance
where
I
is the current in
amperes in a power bus or
an appliance wire and r is
the recommended separation
distance between your device
and the power bus or
appliance wire, in meters (m).
r =
188
(m)
I