EasyManua.ls Logo

Mocacare MOCACuff - Page 51

Mocacare MOCACuff
56 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
49
Appendix
Guidance and manufacturers declaration – electromagnetic immunity
The device is intended for use in the electromagnetic environments
listed below, and should only be used in such environments:
Immunity
test
IEC 60601
test level
Electromagnetic environment – guidance
Conducted
RF
IEC 61000-4-6
Radiated RF
IEC 61000-4-3
3 Vrms 150
kHz to 80
MHz
3 V/m
80 MHz to
2.5 GHz
Portable and mobile RF
communications equipment should
be used no closer to any part of the
device, including cables, than the
recommended separation distance
calculated from the equation applicable
to the frequency of the transmitter.
Recommended separation distance
d = 1.2 √p 80 MHz to 800 MHz
d = 2.3 √p 800 MHz to 2.5 GHz
where P is the maximum output
power rating of the transmitter in
watts (W) according to the transmitter
manufacturer and d is the recommended
separation distance in metres (m).
Field strengths from xed RF
transmitters, as determined by an
electromagnetic site survey,
a
should be
less than the compliance level in each
Not
Applicable
10 V/m
80 MHz to
2.7 GHz
frequency range
b
.
Interference may occur in the
vicinity of equipment marked
with the following symbol:
Compliance
level