BM09201 Molift SMART 150 English - Rev C1/08/10 Page 16 of 20
Maintenance
Maintenance of the lifter consists of check be-
fore use, cleaning, inspection and service. Check
the lifter according to the checklist before use on
page 7, to reveal faults or errors. In regularly in-
spections of the lifter, which as a minimum should
be monthly inspections and annual (periodic) in-
spections.
Cleaning
Detergents must be pH-neutral. Do not use
solvents or strong liquids - this may damage
surfaces on the lifter. For disinfection when
needed; use isopropyl alcohol. Avoid abrasive
cleaning products.
• Cleansurfaceswithadampclothusinganap-
propriate detergent.
• Removehairandpilefromthecastorsandver-
ify that the castors rotates without abnormal
friction.
• Cleancontactandhandcontrolwithspiritor
similar to remove grease and dirt.
• Regularcleaningisrecommendedandshould
be implemented in the suitable cleaning rou-
tines.
Monthly inspection
Assemble the lifter before inspection.
• Checkthelifteraccordingtothechecklistbe-
fore use on page 7.
• Cleanthelifterasdescribedabove.
• Verifythatthecontactsofthehandcontroland
charger is rmly in place, and verify that the ca-
bles for the hand control, actuator and charger
are intact.
• Runthelifterupanddowntoverifythatitruns
normally and without any abnormal noise.
• Theliftermustbetakenoutofserviceimme-
diately and marked “Out of order”, or not taken
into service, if faults and/or irregularities are
discovered.
Periodic inspection
Molift recommends periodic inspection once a
year, alternatively more often if required by local
regulations. Use the checklist for periodic inspec-
tion on page 18.
Inspectionistobeperformedbyauthorizedper-
sonnel. The owner is responsible to ensure prop-
er logging and written verication of each inspec-
tion.