22
2.3.3 Technical Specifications
Environmental Specifications
Temperature of Environment
5 40 °C
Humidity
5 80% non-condensing
Altitude
<2000 m
Power Specifications
Model
VC1110, VC1130
Mains Power Connector
IEC 60320 C14
Line Voltage
100 240 VAC
Frequency
50 60 Hz
Maximum Current Draw
1.9 A
Maximum Power Consumption
190 W
Wall Outlet
N. America & non-EU
Standard NEMA 5-15
CEE Europe (non-UK)
CEE 7/3 or CEE 7/5 w/ common ground terminal
UK
BS1363 (UK) w/ common earth ground terminal
Israel
I-32-3 w/ common earth ground terminal
Physical Dimensions
Component
L x W x H
Mass
VC1110 | VC1130
43 cm x 38 cm x 26.5 cm
24 kg
2.3.4 Safety Information
The following hazards may be typical for this product:
WARNING
Refer to associated product manuals for complete safety information
The VC1130 model contains a Pfeiffer Vacuum® HiPace® 80 turbopump. Refer to the Operating
Instructions provided by Pfeiffer for important safety information regarding this component.
WARNING
Risk of injury due to sharp edges
The interior of the enclosure contains sheet metal parts that may have sharp edges.
When working inside the enclosure (authorized service personnel only), exercise caution to avoid
getting cut by these edges.