Cryostation® s-series
35
3.6 Moving the System
If the system ever needs to be moved to a different lab or location, follow the steps below:
1. Save any important data on the system.
2. Gracefully power down the unit by:
a. Tap (press and release) the power button on the front of the system control unit -or-
b. On the touchscreen UI, navigate to MENU > SYSTEM SETTINGS and select POWER OFF
3. Remove the helium and vacuum hoses, paying special attention to ensure the O-rings remain in
place.
4. Remove the remaining cables and electrical connections.
5. Reverse the steps in Removing the Shipping Supports on page 29. Replace the red locking rings
and c-spacers and lock the red shipping support down on the tri-flange. Then, unbolt the
system from the optical table.
6. Re-pack the system in original packaging. Some components must be packed upright on a
pallet to avoid damage.
NOTICE
Follow all handling instructions for the individual components as outlined in General Hazard
Information on page 10 and the System Overview section starting on page 12.
If you are uncomfortable with moving the system on your own, or if you need to order any
replacement packaging, please contact an authorized service representative.
Do NOT attempt to disassemble any components of the system beyond the original state as
shipped from the Montana Instruments factory.