49
This error only occurs when heat sink temperature (ST
on display or OUTT in Calisto Real Time) is higher than
27°C (81°F) and the external fan is not running. In that
case, the fan is disconnected or has failed. Please contact
Morgan Schaffer for assistance, either by phone at
001.514.739.1967 or by e-mail at
support@morganschaffer.com.
This error occurs when the internal fan is no longer
running. Open the door and manually check the air
flow. If no air circulation is found, the fan must be
replaced. Please contact Morgan Schaffer for assistance.
Please contact Morgan Schaffer for assistance.
Please contact Morgan Schaffer for assistance.
Please contact Morgan Schaffer for assistance.
This error occurs when analog input is not properly
connected or missing.
- Verify the current in the 4-20mA circuit with a
multimeter.
- Verify connections for 4-20mA input pinout; refer
to Figure 9.
- Validate the 4-20mA generator functionality.
If all the above are correct, please contact Morgan
Schaffer for assistance.
This error occurs when the oil flow is below the pre-set
threshold.
- Verify that all valves are open (transformer and
Calisto/Calisto 2).
- Take an oil sample using the Quick Connect Oil
Sampling Tube as described in Section 5.0, and
validate oil viscosity value.
If all the above are correct, please contact Morgan
Schaffer for assistance.
Table 1: Individual Error Code