Battery Information
The rechargeable battery must be charged periodically for at least 24 hours to protect the
battery from fully discharging. If the battery is not charged for an extended period of time it
may lose its ability to hold a charge and will require replacement. The battery may be charged
by plugging the table into AC power or by using the available external battery charger listed in
the Replacement Parts and Upgrade Kits section of this manual.
If the product is being stored for an extended period of time it is recommended the battery be
charged at least 24 hours every 6 months.
The battery should be charged continuously at least 24 hours under the following
circumstances:
First Operation
Before long period storage without AC-in
First operation after long period storage
The battery charge level is indicated by the LED on the battery cradle. The following table
provides an overview of battery status:
Status
LED Indicator
Battery Level
Hi Charging Green Continuous On >80%
Charging Green Blinking (1 sec on / 1 sec off) <80%
Hi Battery Green Blinking (.5 sec on / 4 sec off) >25%
Low Battery Orange Blinking (.5 sec on / 4 sec off) <25%
Protection LED is off <20%
Failure Orange Continuous On -