Cleaning
EchoTable User Manual Page 31 of 45 DCR-00576
(MAN-0011-D)
RECOMMENDED MAXIMUM CLEANER TO WATER SOLUTIONS
Mildest
1:1 mix of mild soap and water. Wipe surface with damp cloth with water
only after cleaning.
Straight application of common disinfectants. Wipe surface with damp
cloth with water only after cleaning.
1:1 mix of ammonia and water. Use a soft cotton cloth saturated with the
cleaning material. Wipe surface with damp cloth with water only after
cleaning.
1:4 mix of bleach and water. Use a soft cotton cloth saturated with the
cleaning material. Wipe surface with damp cloth with water only after
cleaning.
1:1 mix of isopropyl alcohol and water. Use a soft cotton cloth saturated
with the cleaning material. Wipe surface with damp cloth with water only
after cleaning.
Straight application of isopropyl alcohol. Use a soft cotton cloth saturated
with the cleaning material. Wipe surface with damp cloth with water only
after cleaning.
Strongest
1:1 mix of acetone and water. Use a soft cotton cloth saturated with the
cleaning material. Wipe surface with damp cloth with water only after
cleaning.
This information is not a guarantee and does not relieve the user from the responsibility
of proper and safe use of the product and all cleaning agents.
WARNING
It is recommended that the product be cleaned between patients; please follow your facility’s
documented policy.
CAUTION
Always read manufacturer’s instructions and warnings before using any cleaning product or
disinfectant.
Substances such as imaging gels and alcohol will not damage the vinyl surface when
immediately removed. Studies have shown that exposure for longer than a few minutes can
damage the top coat and will eventually discolor vinyl.
Do not use abrasives to clean painted surfaces.