Specifications
EchoTable User Manual Page 35 of 45 DCR-00576
(MAN-0011-D)
Environmental Conditions
ATTRIBUTE Range for Use
Range for Storage and
Transport
Ambient Temperature +0° to 40° C -10° to +50° C
Relative Humidity
20% to 90% @ 30°C
not condensing
20% to 90% @ 30°C
not condensing
Atmospheric Pressure 860 to 1060 hPa 860 to 1060 hPa
Upholstery
ATTRIBUTE
SPECIFICATION
Foam California Technical Bulletin 117
Power Requirements
ATTRIBUTE
SPECIFICATION
Electrical, Product
100–240 VAC, 50/60 Hz, max 5 A
Electrical, Battery Charger
100–240 VAC, 50/60 Hz
Battery Option
24 VDC, 2.9 Ah
Duty Cycle
10%, 2 min. on / 18 min. off
All electrical circuitry is isolated from chassis.
Grounding reliability can only be achieved when the equipment is connected to
an equivalent receptacle marked “Hospital Only” or “Hospital Grade.
The power cord is to be used for mains disconnection.
Attached power supply cord set is not allowed to be diverted to other
equipment.