28 | ni.com | NI 5114 Calibration Procedure
2. Configure the input impedance and the maximum input frequency using the 
niScope Configure Chan Characteristics VI.
3. Configure the common vertical properties using the niScope Configure Vertical VI.
LabVIEW VI
C/C++ Function Call
Call 
niScope_ConfigureChan
Characteristics
 with the 
following parameters:
vi: The instrument handle from
niScope_init 
channelList: "0"
inputImpedance: 
NISCOPE_VAL_50_OHM
maxInputFrequency: The 
Maximum Input Frequency 
value in Table 9 for the current 
iteration
LabVIEW VI C/C++ Function Call
Call niScope_Configure
Vertical
 with the following 
parameters:
coupling: NISCOPE_VAL_DC
probeAttenuation: 1.0
vi: The instrument handle from 
niScope_init
channelList: "0"
range: The Range value in 
Table 9 for the current iteration
offset: 
0.0
enabled: NISCOPE_VAL_TRUE