EasyManua.ls Logo

National Instruments PXI - Adjust Sample Rate; Final Self-Calibration and Session End

National Instruments PXI
43 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
NI 5114 Calibration Procedure | © National Instruments | 39
11. Adjust the sample rate using the niScope Cal Adjust VCXO VI.
Note The 10 MHz stimulus is automatically taken from channel 0.
12. Disconnect or disable all inputs to the device.
13. Self-calibrate the device using niScope Cal Self Calibrate VI.
f. End the session and save the results using the niScope Cal End VI.
LabVIEW VI C/C++ Function Call
Call niScope_Cal
AdjustVCXO
with the
following parameters:
sessionHandle: The
instrument handle from
niScope_CalStart
stimulusFreq:
10,000,000
LabVIEW VI C/C++ Function Call
Call niScope_CalSelf
Calibrate
with the following
parameters:
sessionHandle: The instrument
handle from
niScope_CalStart
channelList: VI_NULL
option: VI_NULL
LabVIEW VI C/C++ Function Call
Call niScope_CalEnd with the
following parameters:
sessionHandle: The instrument
handle from
niScope_CalStart
action:
NISCOPE_VAL_ACTION_STORE

Related product manuals