2-32 | ni.com
Chapter 2 Basic Functions
CF18 Unused
CF19 Current and Power Display on Slave Units for Master-slave Parallel
Operation
Sets whether the current or power on slave units is displayed during master-slave parallel
operation. You can display and specify this parameter only when the RMX programmable power
supply is being used as a slave unit.
If this parameter is set so that the current or power on slave units is not displayed, the ammeter
displays”----.”
CF20 LAN Interface Parameter
Sets whether the LAN interface will be used. By setting it to “oFF,” you can disable the LAN
interface function even when a LAN cable is connected. You cannot set the LAN CONFIG
settings while this parameter is turned off.
Even when this is set to “on,” if the command language is set to legacy (CF21: LGCy), you
cannot use the LAN interface.
This parameter is applied when you turn the POWER switch off and on.
Value Parameter Description
oFF Communication monitor timer is set to off (factory default setting).
1 to 3000 1 s, 3 s, 10s, 30s, 100s, 300s, 1000s, 3000s
Value Parameter Description
— Unused
Value Parameter Description
oFF The current or power of slave units is not displayed (factory
default setting).
on The current or power of slave units is displayed.
Value Parameter Description
oFF The LAN interface will not be used.
on The LAN interface will be used (factory default setting).