22 / 40
22 | Set up and Installation
4.8 Performance
ZEROWIRE® G2 HD Wireless Video System is intended and optimized for use in surgical or procedure rooms.
Use outside of a clinical environment is not recommended.
The following steps will help you to achieve optimal performance of the ZEROWIRE® G2 system:
1. Mount both components at least 5feet (1.5m) from the floor.
2. Ideally, both the transmitter and receiver should be at the same height.
3. For a reliable video link, follow the setup guidelines in section Positioning and Orientation [}22].
4. The transmitter and receiver should be facing each other and be visible to each other in free air space.
5. For non line of sight applications, we recommend that the Transmitter and Receiver be located within
6feet (1.8m) or less from the walls.
4.9 Positioning and Orientation
Due to the shape of the antenna’s signal field, both Transmitter and Receiver units should be installed so they
are aligned vertically and horizontally within
±
10°.
Vertical Alignment
Vertical alignment should be within
±
10°.
Horizontal Alignment
Horizontal alignment should be within
±
10°.
(1) Transmitter
(2) Primary monitor
(3) Receiver
(4) Secondary monitor