5 Operating Instructions
5-2
5.2.2
Parameters of Poring Pulse mode
Voltage
The voltage of the first Poring Pulse (increments: 1V)
Pulse Length
The time duration of one Poring Pulse in milliseconds (increments: 0.1ms)
Pulse Interval
The time interval between Poring Pulses or the last Poring Pulse and the first Transfer Pulse in
milliseconds (increments: 0.1ms)
Number of Pulses
The number of Poring Pulses
Polarity Switching
This function reverses the polarity of Transfer Pulses. When the “+” indicator of the Polarity selection
option in the Set Parameters frame is illuminated, it means that the polarity switching function is OFF.
When the “+/-“indicator of the Polarity selection option is illuminated, it means that the polarity switching
function is ON. If the voltage, number of pulses are programmed, the out put pulses will be as follows;
5.2.3
Setting parameters of Poring Pulse mode
1
Use the
buttons to select the desired parameter from within the Poring Pulse mode Selection
window.
2
When your choice is made: Press the Set button.
Time
Voltage