Operator’s Handbook
Diagnostics
Diagnostics
Errors and Warnings
In the event one of these messages appears on the display, take the action described in the "Solution" section of the message.
Errore 1: H2O flux failure
The flux sensor does not measure the presence of liquid in the cooling system.
Check out the coolant level
Check out and clean up the water filter
Errore 2: Cap. Charge fail
Internal capacitors are not fully charged or not charged at all
• Check if there are problems on the power supply electric network
• Check the actual working voltage
•
The flash lamp is not ignited (Simmering)
•
Errore 4: Shutter failure
The cavity safety fails to close or is malfunctioning
The comunication beetwen the circuit board inside the machine is not working
Warnings
All the below ”Warning” messages are displayed because of external conditions affecting the proper functioning of the equipment and can be
solved without switching it off
Warning 1: Remote interlock
Close the remote interlock contact to start working again
Cooling liquid temperature too high
Wait until the machine re-starts automatically.
(Do not turn the system off!)
If the machine doesn’t start automatically call for the service
Electronics temperature too high
Wait until the machine re-starts automatically.
(Do not turn the system off!)
If the machine doesn’t start automatically call for the service
Do not place the equipment in a hot environment. An increased ambient temperature reduces the efficiency of the cooling system.
Warning 4: Pendrive unplugged
No pen drive in, no data saving nor retrieving. Please check if the pen drive is plugged properly in.
Check for correct insertion of the pen drive.
The machine will work without pen drive but user cannot save settings.