• To unlock the Trial Stimulator, press both the plus and minus buttons at the
same time and hold them down for at least two full seconds. The Trial
Stimulator will beep once for one second to confirm it is unlocked.
• When the Trial Stimulator is unlocked, you can use the plus and minus buttons
to make adjustments to the stimulation strength.
• The Trial Stimulator will lock itself again automatically if you do not press any
button on it for a minute or more.
Plugging OR Cables into the Trial Stimulator
You will receive the Trial Stimulator with OR Cables. A diagram of these cables and the
Trial Stimulator appears in the Appendix. To plug them into the Trial Stimulator, follow
these steps:
1. Make sure the Trial Stimulator is turned OFF by pressing the red Stimulation
OFF Button.
2. Select one of the OR Cables (it does not matter which one you start with).
Check the color of the sticker at the end of the cable that plugs into the Trial
Stimulator.
Patient Manual 11052 Rev A (2015-01-16) 22