EasyManua.ls Logo

Newport HT50 - Page 26

Newport HT50
123 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
DESCRIPTION OF CONTROLS, INDICATORS, ALARMS & CONNECTIONS
3-2 SERHT50NA A0808
On Condition: SPONT
or
Setting Condition: A/CMV, SIMV, or SPONT
Select—Adjust (Up / Down)—Accept
Select the Pressure Control or Volume Control button. Both the
LED indicator and target v
alue will blink.
Adjust the blinking target value for the selected control with the
Up / Down buttons.
Accept the new control and target value by either pressing the
selected b
utton again; or by pressing another button to select a
new parameter for adjustment, or by waiting 5 seconds without
making a change.
Note: The transition to a new pressure or volume target may
require se
veral breaths.
Changing a Parameter (or Multiple Parameters)
Select—Adjust (Up / Down)—Accept
Select the parameter by pressing the labeled button (i.e. ƒ, P trig,
etc).
The parameter’s numeric display will blink.
Adjust the numeric value with the Up / Down buttons.
Accept the value by either pressing the selected button again; or
by pressing another b
utton to select a new parameter for
adjustment, or by waiting 5 seconds without making a change.
Enabling/Disabling Auto Panel Lock
Auto Panel Lock can be enabled or disabled via User Set Up (see
pg 3-27). When the Auto Panel Lock is enabled, the Panel will lock
30 seconds after the last button is pushed and the LED lights
green. All touch buttons (except Silence/Reset and Internal
Battery Test) are locked, preventing accidental parameter
changes.
NOTE: Auto P
anel Lock is factory preset to “Disabled” (off).
To temporarily unlock parameters when Auto Panel Lock is active,
push the Push To unlock button for at least one second. The
Panel will relock 30 seconds after the last button is pushed.

Table of Contents

Related product manuals