DESCRIPTION OF CONTROLS, INDICATORS, ALARMS & CONNECTIONS
3-26 SERHT50NA A0808
External Power Connector
100-240 VAC, max. 2A 12-30 VDC, max. 12A
50/60/400 Hz
The HT50 uses a single inlet for both A.C. and D.C. power sources.
The inlet power connector automatically recognizes A.C. voltage
ranges from 100 to 240 and D.C. ranges from 12 to 30. A Newport
approved external battery can be attached to this connector.
WARNING To ensure proper grounding and prevent possible
shoc
k hazards, this device should only be connected to
“Hospital Grade” power receptacles.
WARNING HOME CAREGIVERS: External power in the
home en
vironment must support min. 100 to max. 240 VAC
and must have a grounded receptacle.
Equipotential Connector
Used for electric potential equalization.
Power Cord Ferrite
Use of the Power Cord Ferrite ensures that the HT50 meets EMC
requirements. Anytime the HT50 is operating on A.C. power, the
ferrite should be attached to the power cord. Operating from D.C.
power does not require the use of the ferrite.
NOTE: Alwa
ys use the A.C. power cord supplied with the HT50.
OPTIONAL ACCESSORIES
Air/Oxygen Entrainment Mixer
The Air/Oxygen Entrainment Mixer (p/n HT460700) is used to
blend atmospheric air with medical grade oxygen at a precise
ratio. A control knob allows for incremental adjustment from 0.21
to 1.00 F
I
O
2
. The high pressure oxygen hose has a standard
female DISS 1240 connection. The Mixer attaches into the Fresh
Gas Intake of the HT50 on the Filter Cover, located on the right
side of the ventilator.
Pneumatic Requirements
Oxygen 35-90 psig (2.4 to 6.2 Bar)
D.C. connections
A.C. connections