SECTION 3
SERHT50NA A0808 3-29
Set Up
The Set Up parameter allows the operator to set the ventilator’s
star
t up settings. There are two selections, DEFAULT or USER.
Use the Up/Down button to toggle between selections.
When USER is selected, all of the v
entilation parameters in effect
at shutdown will be saved. The saved parameters will appear next
time the ventilator is powered on.
When DEFAULT is selected, factory set default parameters will
appear next time the ventilator is powered on. Default parameters
are listed on pg 2-7.
NOTE: HOME CAREGIVERS: This parameter should always be set
to USER f
or home use to ensure that when the HT50 is powered off
and on the physician directed ventilation settings remain in place.
Auto Panel Lock
This setting allows the user to enable (turn on) or disable (tur
n
off) the Auto Panel Lock feature. For a complete description of
Auto Panel Lock see pg 3-13. The Auto Panel Lock feature is
factory set at Disabled or off. Use the sUp/tDown buttons to toggle
between Enabled and Disabled.
NOTE: HOME CAREGIVERS: Newport recommends that the Auto
Panel Lock feature be enabled in homecare environments as an
added safety f
eature to prevent accidental changes to panel controls.
Exiting User Set Up
Exit by pressing the Silence/Reset button at any time.