WARNING The ventilator is a critical life support device. Upon
replacement of any critical assembly, it is important to complete
a 24 hour burn-in to assure proper operation.
1. To complete the burn-in test, set the ventilator to the Standard
Test Settings detailed in Section 5.
2. Let the ventilator operate for 24 hours.
Operational Verification Procedure
Upon completion of the Burn-in test, perform all calibrations and
Operation Verification Procedures as detailed in Section 5.