way. If the product is damaged, return it to NI for
repair.
Safety Voltages
Connect only voltages that are within the following limits.
...........................................................Channel-to-COM ±30 VDC max
Isolation Voltages
...........................................................Channel-to-channel None
Channel-to-earth ground
Continuous
...........................................up to 2,000 m
altitude
250 Vrms, Measurement
Category II
...........................................up to 5,000 m
altitude
60 VDC, Measurement
Category I
Withstand
...........................................up to 2,000 m
altitude
2,300 Vrms, verified by a
5 s withstand test
...........................................up to 5,000 m
altitude
1,000 Vrms, verified by a
5 s withstand test
NI 9203 Getting Started Guide | © National Instruments | 3