Nikon Eclipse E400 POL - Image Quality and Focusing Difficulties; Care and Maintenance

Nikon Eclipse E400 POL
44 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Problem
Cause
Countermeasure
The
image
is
yellowish.
An
NCBI
filter
is
not
being
used.
Use
the
NCBI
filter.
(p.20)
The
lamp
voltage
is
too
low.
The
image
is
too
bright.
The
lamp
voltage
is
too
high.
Inadequate
illumination
(also
check
the
electrical
system
problems
and
countermeasures)
The
lamp
voltage
is
too
low.
Push
auto-photo
switch
and
then
adjust
the
brightness
through
the
ND
filter
combination
(p.20
and
21)
The
aperture
diaphragm
is
stopped
down
1oo
far.
Open
the
diaphragm
to
a
suitable
size.
(p.19)
The
condenser
is
too
low
Position
the
condenser
so
that
the
image
of
the
field
diaphragm
forms
properly
on
the
specimen.
(p.18)
Focusing
is
not
possible
with
high-power
objectives.
The
specimen
is
placed
upside-down.
we
with
the
Set
the
specimen
on
the
cover
glass
facing
up.
The
cover
glass
is
too
thick.
Use
the
specified
type
of
cover
glass
(thickness:
0.17
mm)
The
objective
strikes
the
specimen
when
changing
from
a
low-power
objective
to
a
high-power
objective.
The
specimen
is
placed
upside-down,
Set
the
specimen
on
the
stage
with
the
cover
glass
facing
up.
‘The
cover
glass
is
too
thick
Use
the
specified
type
of
cover
glass
(thickness:
0.17
mm).
The
diopter
setting
has
not
been
adjusted.
Adjust
the
diopter
setting.
(p.17)
The
difference
in
focus
is
large
when
the
objective
is
changed.
The
diopter
setting
has
not
been
adjusted.
Adjust
the
diopter
setting
(p.17)
When
viewing
through
the
binocular
eyepiece,
the
image
does
not
resolve.
into
a
single
image.
The
interpupillary
distance
has
not
been
adjusted.
Adjust
the
interpupillary
distance.
(p.17)
The
diopter
setting
has
not
been
adjusted.
Adjust
the
diopter
setting.
(p.17)
Eye
strain
develops
while
viewing.
The
interpupillary
distance
has
not
been
adjusted.
Adjust
the
interpupillary
distance.
(р.17)
The
diopter
setting
has
not
been
adjusted.
Adjust
the
diopter
setting.
(p.17)
The
brightness
level
is
not
suitable.
Adjust
the
brightness
through
the
ND
filter
combination.
(p.20)
36