15
Glucose Control Test
We recommend performing Control Tests for
practice before using your Meter for the first time
to test your blood.
Control Tests should be performed:
For Practice to ensure your testing technique is good,
Occasionally as you use the vial of Strips,
When opening a new vial of Strips,
If results seem unusually high or low,
If a vial has been left opened or exposed to extreme
heat or cold, or humidity,
If Meter damage is suspected (Meter was dropped,
crushed, wet, etc.)
Note:
Two levels of TRUEcontrol Glucose Control
Solution are available for Quality Control Testing.
It is important to perform Control Tests with more
than one level of Control to assure your System is
working properly and your testing technique is
good. Contact place of purchase or call for
assistance to obtain Control.
Control ranges printed on Test
Strip vial label are for Control Test
results only and are not
suggested
levels for your blood glucose. Do not
drink Glucose Control.
Only the most current Control Test result is
stored in Memory.
TB_OB_GutsENG_H4IAU03.qxd:TT_OB_GutsENG.qxd 10/13/10 12:32 PM Page 15