49
System Specifications
Result Range: 20-600 mg/dL
Sample Size: 0.5 microliter (0.5 µL)
Sample:
F
resh capillary whole blood from the nger and
forearm, or control solution
Test Time: Results in as little as 4 seconds
Result Value: Plasma values
Assay Method: Amperometric
Power Supply: One 3V lithium battery
#CR2032 (non-rechargeable)
Battery Life: Approximately 1000 tests or 1 year
Automatic shut-o: Aer two minutes of non-use
Weight: 1.66 ounces
Size: 3.44” x 2.16” x 0.69”
Memory Size: 500 glucose results
Operating Range (meter & test strips):
Relative Humidity: 10-90% (Non-condensing)
Temperature: 41-104°F
Hematocrit: 20-70%
Altitude: Up to and including 10,200 feet
Note:
Use within specified environmental conditions only.
Chemical Composition
TRUE METRIX Test Strips: Glucose dehydrogenase-FAD
(Aspergillus species), mediators, buers and stabilizers.
TRUE METRIX Control Solution: Water, d-glucose,
buers, viscosity enhancing agent, salts, dye and
preservatives.
TMX_OB_GutsENG_RE4NPD03r6.indd 49 7/15/15 3:37 PM