Changing the Battery
If you have a spare fully charged battery, it can be changed
to allow for continuous operation.
WARNING: Replace the battery with Nova P/N 42215
only. Using another battery may present
a risk of re or explosion. If discarding,
dispose of the battery promptly. Keep the
battery away from children.
1. Press the Power button to enter the Sleep Mode. This will
allow the operator approximately 20 seconds to change
the battery and not lose date/time settings.
NOTE: Ifittakeslongerthan20secondstochangethe
battery,powerupthemeter,relogin,andsetthe