5
EN DE FR ES IT PT
•Holdthespraycanupright.
CAUTION
9-4 Lubrication
NSKPANASPRAYPlus
ApplyNSKPANASPRAYPluseverytimeaftereachuseand/orbefore
autoclaving.
1)Removethemotorfromthehose.
2)Mountthearrow-headspraynozzletipintothespraycanport.
(Fig.7)
3)InserttheSprayNozzleintotheDriveAirPortofthemotor.Holdthe
motorandsprayforapproximately2-3seconds.Applylubricant
untilitexpelsfromtheexhaustairpartforatleast2seconds.(Fig.8)
LubricationOil
1)Removethemotorfromthehose.
2)Apply1-2dropsoflubricationoilintotheDriveAirPortshownin
theFig.8.
3)Connectthemotortothehose.Operatethemotorforabout10
seconds.
9-5 Sterilization
Sterilizetheproductbyautoclavesterilization.Aftereachpatient
sterilizeasnotedbelow.
1)Insertintoanautoclavepouch.Sealthepouch.
2)Autoclavableundertheconditionsbelow.
Autoclaveformorethan20min.at121°C,or15min.at132°C,or
3min.at134°C.
Wipethesurfacewithclothsoakedindisinfectantsolutionsuchas
alcohol.
AutomaticDisinfection(ExternalandInternalSurfaces)
(OnlyforFX204mM4/FX204mB2/FX205mM4/FX205mB2)
Useathermo-disinfectorinaccordancewithISO15883(ENISO15883).
Refertothethermo-disinfectormanual.
•Besuretoholdthehandpiecefirmlytopreventslippingoutof
handbythespraypressurewhenlubricating.
•NSKrecommendstheuseof"SprayMistAbsorber"(Y900084)to
preventoilmistexpellingoutofthemotor.