6
Points to check
Rotation
Details
Rotate the motor and check for abnormalities
such as abnormal rotation, vibration, noise, and
overheating.
11. Symbol
ThisproductisAutoclavableuptoMax.135°C.
10. Periodical Maintenance Checks
Performperiodicalmaintenancecheckseverythreemonths,referring
tothechecksheetbelow.Ifanyabnormalitiesarefound,contactyour
AuthorizedNSKDealer.
3)Thehandpieceshouldremainintheautoclavepouchuntilrequired
foruse.
•Donotautoclavetheproductwithotherinstrumentsevenwhenit
isinapouch.Thisistopreventpossiblediscolorationanddamage
totheproductfromchemicalresidueonotherinstruments.
•Donotheatorcooltheproducttooquickly.Rapidchangein
temperaturecouldcausedamagetotheproduct.
•Toavoidproductfailure,donotuseasterilizerwhosetemperature
exceeds138°Cduringuseincludingthedrycycle.Thechamber
temperatureofsometypeofsterilizersmayexceed138°C.
Contactthesterilizermanufacturerfordetailedinformation.
•Keeptheproductinsuitableatmosphericpressure,temperature,
humidity,ventilation,andsunlight.Theairshouldbefreefrom
dust,saltandsulphur.
•Donottouchtheproductimmediatelyafterautoclavingasitwillbe
veryhotandmustremaininasterilecondition.
CAUTION
•NSKrecommendsClassBsterilizersasstatedinEN13060.
NOTICE
•Autoclavesterilizationisrecommendedfortheproduct.The
validityofothersterilizationmethodsisnotconfirmed.