8
2.6 STANDARD ACCESSORIES
To make the warranty active it is necessary that a copy of the supplied Unit Passport is sent, through
the agent, to the manufacturer; for want of this the warranty will decline.
2.7 TECHNICAL SPECIFICATIONS
2.7.1 Environment operating conditions
The sterilizer is designed to operate in environments with temperature between 3 °C and 40 °C, relative hu-
midity <95%, atmospheric pressure from 750 mBar to 1050 mBar and altitude between 0 and +2000m.
Ø = 240 mm
Depth = 384 mm
elapsed 30’ from the end of a
cycle and without any action
4 liters each (waste and
fresh water tanks)
4 kg (solid)
1,5 kg (porous)
20’ from room temperature
10’ with pre-heated chamber
Clock battery
(may be replaced only by au-
thorized service)
from 3’ to 90’ depending on the
selected cycle
Transmitted heat
in environment
from 6’ to 14’ depending on the
selected cycle
443 x 590 x 428 mm
(L x D x H)
Water conductivity
monitoring
Maximum volume available
on the trays
Maximum chamber
temperature
Safety valve
working pressure
2 x 12,5A T (type 6.3 x 32 CT ) -
IEC 127
NEVER USE IN PRESENCE OF FLAMMABILE OR ANAESTHETIC GAS.
Bacterial filter
(P/N 021008)
(+ 1 pcs extra)
Water filter
(P/N 105320)
(+ 2 pcs extra)
Filter spanner
(P/N 105228)
2+2 Trays
(P/N 105076 - 105077)