23
7. TROUBLESHOOTING
7.1 INITIAL AUTO-TEST
Each time the equipment is switched on, starts an automatic test (duration approx. 15 s) checking sequen-
tially any main component.
Three beeps signal the end of the auto-test and, if positively passed, the message Card Good is displayed .
Whatever fault detected, it will be shown on the display and stored with the alarm code of the Table C (see
chapter ALARM).
To skip the auto-test, turn on the unit and push a key immediately after.
7.1.1 Water quality check
To prevent the accidental use of poor quality demineralized water, the sterilizer is equipped with a device that
monitors the water conductivity; this check starts at the switching-on (initial auto-test), provided that the unit is
at environment temperature and with the main tank full.
At the end of the auto-test, the display will show the message « H2O good » or « H2O hard » if the meas-
ured conductivity value is respectively lower or higher than 15 µS.
ATTENTION
The negative result of the water quality check does not stop the operation of the sterilizer; anyway, it
is strictly recommended to replace the water with a type of better quality.
The following table shows the minimum values suggested for the feeding water.
heavy metal residues, included
iron, cadmium and lead
colourless, clean, without sediment
colourless, clean, without sediment