4.3 RUNNING A STERILIZAZION CYCLE
1. Turn on the equipment by means of the switch located on the rear panel.
The display shows date, time and OFF; pushing UP, DOWN or START for short time) the brightness
of the display increases for few seconds
2. Push the key Power and wait for a few seconds the initial auto-test completion; during this time the pa-
rameter set-points and the type of the components currently tested will appear in a sequence on the dis-
play. Over the auto-test, the display will show the value of the current pressure and chamber temperature
(if lower than 35 °C the display will show the message low”). The microprocessor enables the pre-heat-
ing step in order to rise the chamber temperature up to 100 °C.
The pre-heating phase is designed to keep warm the chamber so the cycles are faster, this function
increases the power consume in stand by condition; if the unit is not continuously used and the high
speed is not required, it is possible to set the unit in ECO MODE that keeps the chamber cold when
not needed. This system has not effect on the performance of the cycle. Look at SETUP pages.
During this phase the temperature reading on the display is inaccurate, because no steam being
there.
3. Arrange the material to be sterilized on the trays, load the chamber and close the door.
4. Check the indication of the water level, if missing fill up the main tank with demineralized water up to the
max level.
4.3.1 Available sterilization cycles
(*) In case of needs related to local regulations, the sterilization time of all the cycles may be increased on request by author-
ized technical service
(*) The cycle not needed may be eliminated from the selection on request by authorized technical service
(*) Liquids excluded
To select the programs push UP or DOWN.
The display will show the parameters selected and the type of admitted load
PROGRAM
LOAD TYPE
MAX LOAD
CYCLE TYPE
N.
NAME
VACUUM
PHASES
TEMPERATURE
(°C)
STERILIZ.
TIME (min.)
DRYING
TIME (min.)
HOLLOW
A
HOLLOW
B
SOLID
WRAPPED
SOLID
(Kg)
PROROUS
(Kg)
1
UNIVERSAL
3
134
5
10
Yes
Yes
Yes
Yes
4
1,5
B
2
DELICATE
3
121
20
12
Yes
Yes
Yes
Yes
4
1,5
B
3
FLASH
2
134
4
5
Yes
Yes
Yes
No
4
No
S
4
SMALL LOAD
3
134
4
5
Yes
Yes
Yes
Yes
0,5
No
B
5
PRION
3
134
18
10
Yes
Yes
Yes
Yes
4
1,5
B
6
CRITICAL 134°
4
134
5
14
Yes
Yes
Yes
Yes
4
1,5
B
7
CRITICAL 121°
4
121
20
16
Yes
Yes
Yes
Yes
4
1,5
B
8
SPECIAL
2, 3 OR 4
105 TO 135
3 TO 90
5 TO 14
DEPENDS ON THE VALUES SELECTED
T
TEST
3
134
3,5
10
ONLY FOR TESTING PROCEDURES
PARAMETERS