Although the sterilizer’s average life is about 8-12 years, it is compulsory to carry out periodic controls in order
to check the correct calibration and for possible wear of components. The control interval, as reported on the
Service Book, is 1 year or 1500 cycles for the ordinary maintenance, and 4 years or 10.000 cycles for the
special maintenance. The sterilizer is configured to signal on the display, as the first term comes to expire,
the message “ NEED SERVICE”. This feature is in compliance with the norm EN13060 and answers to spe-
cific safety requirements. The maintenance activities must be performed by authorized technicians ( provided
with card or certificate released by NSK ) and reported on this Service Book.
Model ................................................................. Serial Number .............................
Installation Date ................................................... Altitude .......................................
Installer ...........................................................................................................................
Reseller ............................................................................................................................
Sterilization Manager: ...................................................................................................... .
Standard periodic checks
- Disk-door clearance check
- Closing pin replacement
- Screw hinge replacement
Filter cleaning / replacement
- Bacterial filter replacement
*) To be carried out in factory’s service department
In case of direct ship or consign for shipping the sterilizer for service center or factory repairing, include a copy of
the filled Service Book pages.