19
5. PROGRAMMING
SETUP MENU
Push together the buttons UP and DOWN to enter in the setup pages, use DOWN (or UP to start from the
last option) to select the function, START to set, POWER to go back.
The following function may be adjusted:
SET TIME the display shows SET TIME”, push START to enter;
the display shows YEAR SET, using UP and DOWN select the year than push START
the display shows MONT SET, using UP and DOWN select the month than push START
the display shows DAY SET, using UP and DOWN select the day than push START
the display shows HOUR SET, using UP and DOWN select the hour than push START
the display shows MIN SET, using UP and DOWN select the minute than push START
The setup of the time is completed and the display goes back to SET TIME, with DOWN (or UP) select
the others menu.
SET SPECIAL CYCLE this is the setup of the cycle 8 SPECIAL, push START to enter;
the display shows SPECIAL TEMPERATURE, using UP and DOWN select the value than push
START
the display shows SPECIAL TIME, using UP and DOWN select the value than push START
the display shows VACUUM - SPECIAL, using UP and DOWN select the value than push START
the display shows TIME - DRY - SPECIAL, using UP and DOWN select the combination of vacuum +
ventilation drying time than push START
The setup of the cycle 8 SPECIAL is completed and the display goes back to SET SPECIAL CYCLE,
with DOWN (or UP) select another menu.
MEMORIES this menu shows the data stored in the unit, these are not adjustable, push START to enter
and view the following info (using DOWN or UP to switch between items):
CYCLES cycles counter
ABORTED CYCLES failed cycles counter
ALARMS last alarms code
CLEANING CYCLES number of cleaning cycles done
INSTALLATION DATE date of the first installation
LAST SERVICE DATE date of the last technical maintenance
Push POWER to go back to “MEMORIES”, with DOWN (or UP) select another menu.
ENERGY in this menu we can select the power mode. Push START to enter, DOWN or UP to switch be-
tween options:
NORM uses the maximum power in the preheating phase to preheat the chamber and minimize the
holding time.