4.4. MAIN SUPPLY
The centrifuge requires 230 V, 50 / 60 Hz. Please make sure that the supplied mains match
the required power ratings. If no, provide an extra line to support.
SECTION 5 – OPERATING UNIT
5.1. OPERATING
• Open the lid and check that there is no sample in the instrument.
• Turn the centrifuges on by using On/Off switch.
• See that the display and control panel activates.
• Learn the functions of the control panel.
• Set the values and start the operation.
5.2. DISPLAY AND CONTROL PANEL
According to the standard IEC 61010-2-020, anyone and any
hazardous materials should not be in the 300 mm safety zone while
centrifuge is running.
The centrifuges are designed to operate safely under the following
conditions:
- Indoor use only
- Ambient temperature: 5
°
C to 25
°
C.
- Maximum relative humidity of 80% for temperature up to 22 °C.
- Maximum altitude: 2000 m.
- Temperature for maximum performance: 15 °C / 25 °C.
Always plug the sterilizer/ovens to properly earthed sockets.
A supply fitted with a circuit breaker should be used for protection
against indirect contact in case of an insulation fault.