EasyManua.ls Logo

NUVE TK 252 - Page 14

Default Icon
26 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Select by using “TAB” key and selected parameter can be changed by using INCREASE
DECREASE” keys between “YES” or “NO”. After setting the parameters;
- Save the changes in memory by pressing “ENTER” key
- or Pass to the next window without saving the changes by pressing “NEXT” key.
- or Pass to the previous window without saving the changes by pressing “BACK” key.
The next window permits to adjust the temperature and humidity alarm periods.
ALARM PERIODS
TEMPERATURE 10 ºC
HUMIDITY 20 %RH
EXIT PID
Select by using “TAB” key and selected parameter can be changed by using INCREASE
DECREASE” keys. After setting the parameters;
- Save the changes in memory by pressing “ENTER” key.
- or Pass to the next window without saving the changes by pressing “NEXT” key.
- or Pass to the previous window without saving the changes by pressing “BACK” key.
The temperature alarm period can be adjusted between 2ºC and 10ºC. The humidity alarm
period can be adjusted between 5% and 20% RH. (For example temperature alarm period is
chosen 2ºC and if the temperature exceeds ±2ºC, the alarm will be seen on the display. ‘!’ will
blink in front of temperature value.)
The next window permits to adjust the temperature and humidity values.
1 STEP:
TEMPERATURE
040.6 ºC
HUMIDITY
78 %RH
EXIT
STEP NO” can be adjusted between 1 and 9, “TEMPERATURE” can be adjusted between -
20ºC and 60ºC, and HUMIDITY” can be adjusted between 20% RH and 95% RH. The
parameters are selected by using TAB” key and selected parameter can be changed by using
INCREASE–DECREASE” keys. After setting the parameters;
- Save the changes in memory by pressing “ENTER” key
- or Pass to the next window without saving the changes by pressing “NEXT” key.
- or Pass to the previous window without saving the changes by pressing “BACK” key.
13

Related product manuals