Cooling failure: While the program is running, if the compressor is failed or there is a
problem about the cooling system !warning appears beside the temperature and humidity
value. In this case you press “F4” key that corresponds to “HELP” expression to see the
explanation of the error.
COOLING
SYSTEM FAILURE
CALL SERVICE
EXIT
The window turns back to stand-by position if you wait for 15 seconds or press “F4” key that
corresponds to “EXIT” expression.
Communication error: If there is a problem about the communication between the display
and main PCB, “COMMUNICATION ERROR” appears on the display.
COMMUNICATION
ERROR
The window turns back to stand-by position as soon as the problem is overcome.
Power failure: If the setting ‘IS PROGRAM AFFECTED BY POWER FAILURE?’ is set as
‘YES’ during programming, the program is cancelled because of the power failure and
POWER FAILURE” warning appears on the display after the electricity comes back.
POWER
FAILURE
PROGRAM OFF
If the “STOP” key is pressed the window turns to stand-by position.
Broken sensor: In any position if the temperature sensor endings are broken or loose “SBR”
appears on relevant sensor’s value. In this case you press F4 key that corresponds to HELP
expression to see the explanation of the error.
BROKEN
SENSOR
CALL SERVICE
EXIT