e700 User Manual
TABLE 6 - Default Ventilation Setting- CPR
PARAMETER RANGE
DEFAULT
INFANT CHILD ADULT
Tidal Volume (50 - 1400 ml) 100 250 500
PCV (OFF, 4-50 cmH
2
O) OFF OFF OFF
Mv max (2 -40 L) 30 30 30
Mv min (0.5 – 35 L) 0.5 0.5 0.5
P max. (10 -80 cmH
2
O) 40 40 60
P min.
(0-20 cmH
2
O
during I time only)
3 3 3
CPR FOR MASKED PATIENTS
Figure 14.a - CPR waveform for masked patient
On screen chest compressing animation On screen ventilation animation
patient size selection when switching to CPR mode but could be
adjusted to desired values. Optional pressure controlled ventilation
is provided by setting the PCV pressure parameter. If PCV parameter
is selected, the fl ow controlled ventilation will be disabled. The FiO2
is fi xed at 100% oxygen during CPR mode.